Copyright 2016 © BENZO Energy technology Co.,Ltd . All Rights Reserved | Web Design—Tiandiin |
No. |
Item |
General Parameter |
Remark |
|
1 |
Rated Capacity |
Typical
|
1850mAh |
Standard discharge(0.2C C5A) after Standard charge
|
Minimum |
1800mAh |
|||
2 |
Nominal Voltage |
7.4V |
Mean Operation Voltage |
|
3 |
Voltage at end of Discharge |
5.8V |
Discharge Cut-off Voltage |
|
4 |
Charging Voltage
|
8.4±0.03V |
|
|
5 |
Internal Impedance
|
≤160mΩ |
Internal resistance measured at AC 1KHZ after 50% charge
The measure must uses the new batteries that within one week after shipment and cycles no more than 5 times
|
|
6 |
Standard charge |
Constant Current 0.2C5A Constant Voltage 8.4V 0.01 C5A cut-off
|
Charge time : Approx 6h
|
|
7 |
Standard discharge
|
Constant current 0.2 C5A end voltage5.8V
|
|
|
8 |
Fast charge
|
Constant Current 1.0C5A Constant Voltage 8.4V 0.01 C5A cut-off
|
Charge time : Approx 2.5h |